Skip to content

How to test a motor capacitor with a multimeter

Test motor capacitor capacitance out of circuit, interpret OL or low readings and calculate tolerance limits. Allow for meter accuracy near a limit.

CBB65 metal-can capacitor with its label and terminal groups visible.

Compare an out-of-circuit capacitance measurement with the part’s nominal value and tolerance under the specified test conditions. For 20 µF ±5%, the calculated limits are 19–21 µF. A reading near a limit also needs the meter’s accuracy and the agreed acceptance rule.

Measure with the equipment isolated

This work belongs to a qualified technician. Isolate the equipment, prevent re-energization and follow its approved discharge procedure. Verify the absence of hazardous voltage with a suitable instrument before disconnecting the capacitor for measurement. Stored charge can remain after the supply is removed.

Use capacitance mode on a meter with a suitable range and known accuracy. Measuring the isolated component avoids parallel circuit paths affecting the reading. Follow the meter instructions for lead compensation, settling time and any connection requirements. An LCR meter used for acceptance testing also needs the specified test frequency and signal conditions. Fluke: measuring capacitance

Calculate the band from the nominal value

Multiply the nominal capacitance by the tolerance percentage to obtain the allowed deviation. Subtract and add that deviation for the lower and upper limits. The examples below are calculations from nominal labels; they are not measured sample or batch results.

Calculated ±5% capacitance limits
NominalLower limitUpper limit
4.5 µF4.275 µF4.725 µF
20 µF19 µF21 µF
25 µF23.75 µF26.25 µF
35 µF33.25 µF36.75 µF
Close view of a CBB65 label marked 20 µF ±5%, 450 VAC and 50/60 Hz.

20 µF ±5% on the original label

20 µF ±5% gives calculated limits of 19 and 21 µF.

A borderline display needs an accuracy allowance

Suppose a 20 µF ±5% component displays 19.0 µF. That is the calculated lower boundary, but the display is not an exact value. Meter accuracy, resolution and the agreed measurement uncertainty determine whether the result supports acceptance. Use the inspection decision rule rather than rounding the reading into the permitted band.

OL depends on the meter mode and range. It can indicate an over-range or other condition defined in the instrument manual; it does not by itself prove an open capacitor. A zero or unexpectedly low reading also needs the range, connections and measurement setup resolved before classifying the fault.

Record enough information to reproduce the decision

A useful test record names the model, sample or lot, nominal value, tolerance, measured result, method and date. Add instrument identification, calibration status and test conditions. For a dual-run capacitor, record each section against its assigned terminal pair and its own nominal value.

Capacitance alone does not establish insulation resistance, dielectric strength, dissipation factor or endurance. A damaged case or leaking part also needs removal from service and investigation regardless of one capacitance reading. Additional electrical tests require the correct equipment, model limits and safe procedure.

Define the inspection for your order

Include the model, required tests and acceptance conditions with the report format you need.